The Effect of Heat Due to High DC Current Bias on the Resistance Changes of Standard Oil Resistor

Main Article Content

Ploybussara Gomasang
Montisra Tubnium
Warunya KonKan
Supisara Pilawong
Jutarat Tanarom

Abstract

       This study investigates the effect of heat on the resistance change due to high DC current bias in the range of 3 A to 10 A through a 100 mohm standard oil resistor (Model Rubicon 99563). In the calibration
process of high DC current source, high DC current is biased through a standard resistor. It is found the increase in the resistor’s temperature measured by an automatic temperature measurement with a J-type
thermocouple, resulting in the resistance increase. However, the oil can significantly reduce the increasing rate of temperature, and the resistance change becomes stable after applying current through the resistor for up to 265 seconds. This study can be a guideline to improve the calibration process of the high DC current source.

Article Details

How to Cite
[1]
P. Gomasang, M. Tubnium, W. KonKan, S. Pilawong, and J. Tanarom, “The Effect of Heat Due to High DC Current Bias on the Resistance Changes of Standard Oil Resistor”, TEEJ, vol. 5, no. 2, pp. 26–31, Aug. 2025.
Section
Research article

References

ธมลวรรณ นัททีทอง, อลิษา ดาวโรยรัมย์, หฤทัย เพียรศึกษา, จุฑารัตน์ ทานะรมณ์, และ พลอยบุศรา โกมาสังข์, “การยืนยันผลการวัดค่าความต้านทานของตัวต้านทานมาตรฐานย่านต่ากว่า 1โอห์ม ตามค่ากระแสไฟฟ้าจาเพาะด้วยวิธีโวลต์-แอมแปร์,” การประชุมวิชาการงานวิจัยและพัฒนาเชิงประยุกต์ครั้งที่ 15 (ECTI-CARD 2023), 2566.

สำนักงานมาตรฐานผลิตภัณฑ์อุตสาหกรรม, กระทรวงอุตสากรรม,ข้อตกลงร่วมการประชุมเชิงปฎิบัติการ ข้อตกลงร่วม 4007-2567, แนวทางการสอบเทียบแหล่งกาเนิดกระแสไฟฟ้ากระแสตรงโดยวิธีโวลต์-แอมแปร์, ประกาศในราชกิจจานุเบกษา ฉบับประกาศและงานทั่วไป เล่ม 141 ตอนพิเศษ 340 ง, ISBN (e-book) 978-616-617-387-1, 12 ธันวาคม 2567

T. Sanuksan, J. Tanarom and P. Gomasang, "Automated Voltage-Ratio Measurement for Resistance Calibration," 2024 12th International Electrical Engineering Congress (iEECON), Pattaya, Thailand, 2024, pp. 1-4.

P. Weßkamp and J. Melbert, “High-accuracy current measurement with low-cost shunts by means of dynamic error correction,” J. Sens. Sens. Syst., 5, 389–400, 2016.

O. Mironenko, G. Ejzak, W. Kempton, “Integrated Electric Vehicle Shunt Current Sensing System for Concurrent Revenue Metering and Detection of DC Injection,” Energies, 14, 1193, 2021.

D. Kumpanya, S. Thaipanit, and C. Angthong, “Identification of Fractional-order DC Motor Model by Bat Algorithm”, TEEJ, vol. 5, no. 1, pp. 1–8, Apr. 2025.

S. Duangsuwan, W. Deangkwanthong, and P. Inchan, “Development of Smart Electric Energy Consumption Metering in Building with IoT”, TEEJ, vol. 5, no. 1, pp. 9–14, Apr. 2025.

P. Gomasang and J. Tanarom, “A Study of Thermal Effect on High-Current Standard Resistor,” 20th Int. Conf. on Electrical Engineering/Electronics, Computer, Telecommunications and Information Technology (ECTI-CON), 2023.

ธมลวรรณ นัททีทอง, จุฑารัตน์ ทานะรมณ์, พรรณวษา เหล่าสิม, อภิสรา อรุณสันติโรจน์, ชลธิชา อินริน และ พลอยบุศรา โกมาสังข์, “ผลกระทบของการจ่ายกระแสไฟฟ้ากระแสตรงค่าสูงต่อการเปลี่ยนแปลงค่าอุณหภูมิและความต้านทานมาตรฐานในกระบวนการสอบเทียบแหล่งกาเนิดกระแสไฟฟ้ากระแสตรง,” การประชุมวิชาการทางวิศวกรรมไฟฟ้า ครั้งที่ 47 (EECON-47), 2567.

FLUKE Calibration., “5730A High Performance Multifunction Calibrator,” 2012.

FLUKE Calibration., “5725A Amplifier,” 2012.

FLUKE Corporation, “8588A reference Multimeter Product Specifications,” March 2019 Rev. G, 8/22.Y.

Test Equipment Center, Inc., 2025. Leeds-Northrup-4222B-Specifications-4E86D, Available: https://testequipment.center/Products/Leeds--Northrup-4222B, 14 Mar. 2025.

C. Lothongkam, “ Applications of Silicone Insulation in Electrical Engineering”, TEEJ, vol. 4, no. 2, pp. 14–18, Aug. 2024.