INFLUENCE OF FRESHNESS OF EDTA BLOOD ON RELIABILITY OF THALASSEMIA SCREENING TESTS
DOI:
https://doi.org/10.55766/sujst-2023-02-e02352Keywords:
Mean corpuscular hemoglobin, Mean corpuscular volume, Red cell parameters, Thalassemia carriers, Thalassemia screenAbstract
Thalassemia screening tests include one tube (0.36% NaCl) osmotic fragility test (OFT), mean corpuscular volume (MCV), and mean corpuscular hemoglobin (MCH). Reliability of these parameters in stored EDTA blood is the important issue to be concerned. To address this issue, we analyzed the results of these thalassemia screening tests in 50 fresh and stored EDTA blood samples of collected in the Chiang Mai province of northern Thailand. These blood samples; 25 with MCV < 80 fl and 25 with MCV ≥ 80 fl, were stored at room temperature (RT) (approximately 30°C) and in cold (approximately 4°C). OFT, MCV, and MCH were tested in both groups of blood samples everyday for 7 days. It was found that falsely negative OFT results were seen after 1 day and 4 days of storage at RT and in cold, respectively. For MCV, acceptable values were obtained at 0 day and 2-3 days storage at RT and in cold, respectively. No change of MCH values was observed throughout 7-day storage at both RT and cold. It was concluded that when performing thalassemia screening tests in Thailand, fresh EDTA blood must be used for OFT test and MCV determination. However, although MCH seemed to be the most stable parameters, thalassemia screen cannot rely only on MCH value. Therefore, thalassemia screen using these three tests must be done within blood collection day if no cooling facility is available. This information would also be useful for countries in tropical region where ambient temperature is high.
References
Allen, A., Perera, S., Perera, L., Rodrigo, R., Mettananda, S., Matope, A., Silva, I., Hameed, N., Fisher, C.A., Olivieri, N., Weatherall, D.J., Allen, S., Premawardhena, A. (2019). A"One-Stop" screening protocol for haemoglobinopathy traits and iron deficiency in Sri Lanka. Front Mol Biosci., 6:66. https://doi.org/10.3389/fmolb.2019.00066
Antwi-Baffour, S., Quao, E., Kyeremeh, R., Mahmood, S.A. (2013). Prolong storage of blood in EDTA has an effect on the morphology and osmotic fragility of erythrocytes. Inte. J. Biomed Sci and Eng., 1:20-23. https://doi.org/10.11648/j.ijbse.20130102.11
Arif, S.H., Yadav, N., Rehman, S., Mehdi, G. (2017). Study of hemolysis during storage of blood in the blood bank of a tertiary health care centre. Indian J. Hematol Blood Transfus., 33:598-602. https://doi.org/10.1007/s12288-016-0769-5
Chow, J., Phelan, L., Bain, B.J. (2005). Evaluation of single-tube osmotic fragility as a screening test for thalassemia. Am J. Hematol., 79:198-201. https://doi.org/10.1002/ajh.20387
Gulati, G.L., Hylabd, L.J., HKocher, W., Schwarting, R. (2002). Changes in automated complete blood cell count and differential leukocyte count results induced by storage of blood at room temperature. Arch Pathol Lab Med., 126:336-342. https://doi.org/10.5858/2002-126-0336-CIACBC
Huang, S., Hou, H.W., Kanias, T., Sertorio, J.T., Chen, H., Sinchar, D., Gladwin, M.T., Han, J. (2015). Towards microfluidic-based depletion of stiff and fragile human red cells that accumulate during blood storage. Lab Chip., 15:448-458. https://doi.org/10.1039/C4LC00768A
Karimi, M., and Rasekhi, A.R. (2002). Efficiency of premarital screening of beta-thalassemia trait using MCH rather than MCV in the population of Fars Province, Iran. Haematologia., 32:129-133. https://doi.org/10.1163/156855902320387961
Mane, M., Rao, V., Reddy, E.P., and Vaithilingam, A. (2017). Prolonged storage-induced changes in haematology parameters and stability at room temperature for counting red and white blood cells and platelets. Paripex - Indian J. Res., 6:79-96.
Penman, B.S., Gupta, S., and Weatherall, D.J. (2015). Epistasis and the sensitivity of phenotypic screens for beta thalassaemia. Br J Haematol., 169:117-128. https://doi.org/10.1111/bjh.13241
Pranpanus, S., Sirichotiyakul, S., Srisupundit, K., and Tongsong, T. (2009). Sensitivity and specificity of mean corpuscular hemoglobin (MCH): for screening alpha-thalassemia-1 trait and beta-thalassemia trait. J Med Assoc Thai., 92:739-743.
Sirichotiyakul, S., Maneerat, J., Sa-nguansermsri, T., Dhananjayanonda, P., and Tongsong, T., (2005). Sensitivity and specificity of mean corpuscular volume testing for screening for alpha-thalassemia-1 and beta-thalassemia traits. J. Obstet Gynaecol Res., 31:198-201. https://doi.org/10.1111/j.1447-0756.2005.00280.x
Sysmex Corporation, (1998). KX-21 Automated Hematology Analyzer Operator's Manual (Kobe, Japan).
Tatu, T., and Kasinrerk, W. (2012). A novel test tube method of screening for hemoglobin E. Int J. Lab Hematol., 34;59-64. https://doi.org/10.1111/j.1751-553X.2011.01357.x
Tatu, T., and Sweatman, D. (2018). Hemolysis area: A new parameter of erythrocyte osmotic fragility for screening of thalassemia trait. J Lab Physicians., 10:214-220. https://doi.org/10.4103/JLP.JLP_136_17
Winichagoon, P., Thitivichianlert, A., Lebnak, T., Piankijagum, A., and Fucharoen, S. (2002). Screening for the carriers of thalassemias and abnormal hemoglobins at the community level. Southeast Asian J Trop Med Public Health., 33(2):145-150.
Wiwanitkit, V., Suwansaksri, J., and Paritpokee, N. (2002). Combined one-tube osmotic fragility (OF) test and dichlorophenol- indolphenol (DCIP) test screening for hemoglobin disorders, an experience in 213 Thai pregnant women. Clin Lab., 48:525-528.